Parameters |
33: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PCT, MPV, PDW, P-LCR, NE, NE%, LY, LY%, MO, MO%, EO, EO%, BA, BA%, P-LCC*, Mentzer Index*, RDWI*, IG%*, IG#*, Band%*, Band#*, Seg%*, Seg#* (*) RUO parameters (Research Use Only) |
---|---|
Reagents |
5: 2 hemolyzing reagents, 1 diluent, 2 detergents |
Sample volume |
CBC : 25 μL CBC+DIFF : 40 μL Pre-dilution mode : 20 μL |
Throughput |
Up to 90 samples /hour |
Precision (Reproducibility) | WBC : 2.0% or less (WBC : 4.00 × 103/μL or more) RBC : 1.5% or less (RBC : 4.00 × 106/μL or more) HGB : 1.5% or less HCT : 1.5% or less MCV : 1.0% or less PLT : 4.0% or less (PLT : 100 × 103/μL or more) NE% : 5.0% or less (NE% : 30.0% or more AND WBC : 4.00 × 103/μL or more) LY% : 5.0% or less (LY% : 15.0% or more AND WBC : 4.00 × 103/μL or more) MO% : 12.0% or less (MO% : 5.0% or more AND WBC : 4.00 × 103/μL or more) EO% : 20.0% or less OR within ±1.0 EO% (WBC : 4.00 × 103/μL or more) BA% : 30.0% or less OR within ±1.0 BA% (WBC : 4.00 × 103/μL or more) NE : 8.0% or less (NE : 1.20 × 103/μL or more) LY : 8.0% or less (LY : 0.60 × 103/μL or more) MO : 20.0% or less (MO : 0.20 × 103/μL or more) EO : 25.0% or less OR within ±0.10 × 103/μL (WBC : 4.00 × 103/μL or more) BA : 30.0% or less OR within ±0.10 × 103/μL (WBC : 4.00 × 103/μL or more) (Specifications above applies to normal mode) |
Complete QC programs |
L & J / Westgard Multi rule / X bar B(X bar Batch) management |
Options | Software kit MEK-9100, QS-023W Extra sample racks, T411A Waste sensor, JW-910W |
Dimensions and Weight | 675 W × 589 D × 576 H mm, 66 kg |